The First Maintenance Free FT-IR/FT-NIR Analyzer

ABB capabilities encompass one of the largest portfolios in the world for laboratory, at-line and process FT-IR/FT-NIR analyzers.

ABB’s FT-IR/FT-NIR contains fixed components that do not wear out with use, meaning no components to replace or adjust.  With ABB’s FT-IR/FT-NIR, the HeNe LASER is replaced by a solid state laser.  All optics are non-hygroscopic. Purging is not required for protection of optics.

FT-IR MB3000

Versatile Spectrometer

for Solid, Liquid, and Gas

Spectral range from 485 to 8500 cm-1

Resolution better than 0,7 cm-1

Source: Ceramic Globar, 10-year expected life

FT-NIR MB3600

High-Performance Spectrometer

Designed for QA/QC

Spectral range from 2700 to 14980 cm-1

Resolution 0,5 cm-1

Source: Quartz halogen, 10-year expected life

Application

MB3000-CH90

Laboratory gas analyzer

MBGAS-3000CH

Process gas analyzer

MB3000-PH

Pharmaceutical industries

MB3600-CH40

Solid analyzer

MB3600-CH80

PET pack analyzer

MB36000

Industrial QA/QC

MB3600-PH

Lifescience and pharma

MB3600-CH30

Biodiesel application

MB3600-HP10

Hydrocarbon application

MB3600-CH10

Oleochemical application

Spectroscopic Performance

  • Spectral range 485 to 8,500 cm-1
  • Resolution better than 0.7 cm-1
  • Apodized resolution adjustable 1 to 64 cm-1, in increments of 2
  • Maximum signal-to-noise ratio (root-mean-square, 60s, 4 cm-1, at peak response): 50,000: 1
  • Signal sampling: 24-bit ADC
  • Short-term stability: < 0.09 %
  • Temperature stability: < 1 % per C˚
  • Frequency repeatability (@ 1918 cm-1): < 0.001 cm-1
  • Frequency accuracy (@ 1918 cm-1): < 0.06 cm-1
  • Spectral range with DTGS detector 3,700 to 15,000 cm-1
  • Spectral range with InAs detector 3,700 to 12,000 cm-1
  • Spectral range with InGaAs detector 3,900 to 11,000 cm-1
  • Resolution better than 0.7 cm-1
  • Apodized resolution adjustable 1 cm-1 to 64 cm-1, in increments of 2
  • Signal-to-noise ratio (root-mean-square, 60s, 16 cm-1, at peak response) with DTGS detector: > 100 000: 1
  • Signal-to-noise ratio (root-mean-square, 60s, 16 cm-1, at peak response) with InAs detector: > 500 000: 1
  • Signal-to-noise ratio (root-mean-square, 60s, 16 cm-1, at peak response) with InGaAs detector: > 600 000: 1
  • Signal sampling: 24-bit ADC
  • Short-term stability (@ 8000 cm-1) :< 0.09 %
  • Temperature stability (@ 8000 cm-1): < 1 % per ˚C
  • Frequency repeatability (@ 7300 cm-1): < 0.006 cm-1
  • Frequency accuracy (@ 7300 cm-1): < 0.06 cm-1
  • Absorbance reproducibility (toluene): < 0.002 AU